Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 205
 
Calibration Procedures
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
185
To program the sample sequence shown above, follow this flow chart:
 SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30 
 
PREV NEXT MODE SET 
EXIT
SAMPLE RANGE = 500.0 PPB NOX=X.X 
 
 < TST TST > CAL CALZ CZLS 
SETUP
SETUP X.X
PRIMARY SETUP MENU
 
 CFG ACAL DAS RNGE PASS CLK MORE EXIT 
 SETUP X.X SEQ 1) DISABLED 
 
 NEXT MODE 
EXIT
 SETUP X.X SEQ 2) DISABLED 
 
PREV NEXT MODE 
EXIT
 SETUP X.X MODE: DISABLED 
 
 NEXT 
ENTR EXIT
 SETUP X.X MODE: ZERO–HI 
 
PREV NEXT 
ENTR EXIT
 SETUP X.X SEQ 2) ZERO–HI, 1:00:00 
 
PREV NEXT MODE SET 
EXIT
 SETUP X.X STARTING DATE: 01–JAN–02 
 
 <SET SET> EDIT 
EXIT
 SETUP X.X STARTING DATE: 01–JAN–02 
 
  0 4 SEP 0 3 
ENTR EXIT
Toggle to
set day,
month &
year: DD-
MON-Y
Y
 SETUP X.X STARTING DATE: 04–SEP–03 
 
 <SET SET> EDIT 
EXIT
Default
value
is ON
 SETUP C.4 STARTING DATE: 04–SEP–03 
 
 <SET SET> EDIT 
EXIT
Toggle to set
time: HH:MM.
This is a 24 hr
clock. PM
hours are 13-
Toggle
numbers to
set
number of
days
between
procedures
(1 367)
Toggle keys
to set
delay time for 
each iteration 
of the
sequence:
HH:MM
(0 – 24:00)
 SETUP C.4 DELTA DAYS: 1 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4 DELTA DAYS: 1 
 
  0 0 2  
ENTR EXIT
 SETUP C.4 DELTA DAYS:2 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4 DELTA TIME00:00 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4  DELTA TIME: 00:00 
 
  0 0 :3 0  
ENTR EXIT
 SETUP C.4 DELTA TIME:00:30 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4 STARTING TIME:14:15 
 
 <SET SET> EDIT 
EXIT
EXIT returns
to the SETUP
Menu
 SETUP C.4  DURATION:15.0 MINUTES 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4 DURATION 15.0MINUTES 
 
  3 0 .0 
ENTR EXIT
 SETUP C.4  DURATION:30.0 MINUTES 
 
 <SET SET> EDIT 
EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
 SETUP C.4  CALIBRATE: OFF 
 
 <SET SET> EDIT 
EXIT
 SETUP C.4 CALIBRATE: OFF 
 
ON
ENTR EXIT
 SETUP C.4  CALIBRATE: ON 
 
 <SET SET> EDIT 
EXIT
Toggle key
between
Off and
ON
 SETUP X.X TIMER ENABLE: ON 
 
 SET> EDIT 
EXIT
 SETUP C.4 STARTING TIME:00:00 
 
 <SET SET> EDIT 
EXIT
Toggle NEXT button until ...