Section 3: Wiring

04/12

30

51173 Rev. 8

Installing Load Cells, Continued

10. Set the Switches.

Dip (Address) Switch Setup, Smart Sectional Controllers (SSC)

 In each of the smart SSC boxes there is a 10-position dip switch labeled

S1

.

 This switch is used to identify the section in a binary code.
 The switches must be set properly for the scale to operate.

Note:

Switches 1, 2, and 3 are

always OFF.

Switch 5 ON = 700/1000 ohm load cells

OFF = 350 ohm load cells

 Switches 6 thru 10 are used to set the section (section address) numbers.

11. Set the section number according to the following chart.

Section Number

6

7

8

9

10

Section 1

On

Off

Off

Off

Off

Section 2

Off

On

Off

Off

Off

Section 3

On

On

Off

Off

Off

Section 4

Off

Off

On

Off

Off

Switch Settings

12. Continue in this manner until each Smart Sectional Controller (SSC) box has a

unique section number entered on the dip switches.