Gentec-EO MACH 6 User Manual

Page 36

Advertising
background image

Mach 6 User’s Manual Revision 2.1

36



The Controls are:

Measured Average Power

Enter the measured average power from your Power Meter Here. Mach 6 will use this to
compute a new calibration factor for the wavelength in use and store it in the probe in
use.

Read Average Power From Mach 6

Arms the instrument for a data gathering run, takes the measurement batch and then
extracts the Energy and Frequency to use in calculating the average power.


Calculate Probe Response and Send to Probe

Reads the current probe wavelength response curve, calculates the new response point
from the data, then writes the new calculated Probe Response to the Probe memory.

After the probe EPROM is written, Mach 6 will be reset to ensure the new data is
correctly read from the probe.

Load a Default Probe Response

Write a selectable default Probe Response to the Probe memory.