Mach 6 User’s Manual Revision 2.1

8


The software installation is now complete. You may now run the application software.
The software will wait until it finds a Mach 6 instrument on the USB before it becomes
functional. If the Mach 6 is disconnected from the USB or turned off, the software will
return to the wait state until the Mach 6 is on the USB again.

Power Source


The Mach 6 system is supplied with a 10V, 1.5A universal power module. It will accept
voltages from +9V to +15V provided they supply at least 1A and the plug is 5.5mm outer
diameter, 2.1 mm inner diameter, and 11mm in length, with the center being positive.
Plug the supply into the DC power jack located on the front panel.

Probe Connection


The Mach 6 probes use a 7 pin locking connector. See figure 2. Plug in the probe by
lining up the red reference dots and pushing the connector until it latches into its locked
position. To remove the probe, power down the unit and pull on the outer sleeve. The
connector will unlock and disengage.

Indicator LED’s


There are 6 LED’s on the Mach 6 front panel as shown in figure 1.

The Power LED is illuminated when ever an external supply is plugged into the DC
power jack.

The Trigger LED illuminated when the Mach 6 is triggered. This LED functions at all
times, whether the Mach 6 is armed or not.

The Over temp LED illuminates when the measure probe temperature exceeds 65°C. This
LED is only updated when the unit is armed and triggered.