Set measure” options panel, Easure panel – Gentec-EO PC-MAESTRO (Old Software for MAESTRO) User Manual

Page 18

Advertising
background image

PC-

MAESTRO User’s Manual Revision 1.0

17

4.3.1

”Set Measure” Options Panel

Fig. 4-1 The Set Measure panel

Wavelength: Enter the desired wavelength. Press Enter or Tab to make the change effective.

Range: Use the dropdown menu to select the desired range. The chosen value will be the
maximum that the PC-MAESTRO will measure before displaying

OUT”. By default, the

MAESTRO is in

Auto” mode. For more information about the Auto Mode, please refer to the

MAESTRO user manual.

Measure Mode: Use the dropdown menu to set the measurement mode (Power, Single Shot

Energy, Moving Average, etc.). Please note that it can take up to 3 seconds for the PC-

MAESTRO to switch between measurement modes.

Anticipation ON/OFF” and “Attenuator

ON/OFF

” can be controlled in this menu. Please refer to the MAESTRO user manual to get

more information about these functions.

Corrections: Use these settings to set a multiplier and/or an offset to your measurement. The

equation used is : Result = (Value*Multiplier)+Offset

Trigger Level:

Adjust the trigger level in 0.1% steps, from 0.1% to 99.9%.

Trigger level is not available in power mode

External trigger option is not available in power mode

Default value is 2%

For more information about the trigger level functionality, please refer to

the MAESTRO user manual.