Vanguard CT-8000 S3 User Manual

Page 26

Advertising
background image

CT-8000 S3 USER’S MANUAL

REV 1

21

3.1.9. Resistor Type Transducer Connection

The CT-8000 S3 provides three input channels for resistor type transducers. The transducer
resistance input can range from 200 Ohms to 10K Ohms. The CT-8000 S3 supplies a 5 Vdc
reference voltage to power the resistor type transducer. The sense voltage is translated into a
travel distance based on the transducer calibration parameters. The user is required to setup
the resistor transducer before performing a test. Up to nine transducer setups can be stored in
the CT-8000 S3’s internal memory. Once a transducer setup is stored, it can be recalled quickly
before running a test. A typical resistor type transducer connection is shown in Figure 13.

NOTES

• When a resistor type transducer is selected, all three resistor transducer channels

(R1, R2, and R3) are active and will be shown on the test results (see Figure 18).

• A 15-foot transducer cable with 3 quick disconnect clips is provided with each CT-

8000 S3. The 3 clips are labeled “+5V”, “Signal”, and “GND”.

• See section 3.2.11 for resistor type transducer calibration information.

Figure 13. Typical Resistor Type Transducer Connection

Advertising