Operating controls and indicators, power, charger on – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual

Page 12: batt. ready (quickcharge), fault, testing, test button, Setup, Battery charging procedure (quickcharge), 2shudwlqj &rqwurov dqg ,qglfdwruv

Advertising
background image

ZOLL Base PowerCharger

4x4

Operator’s Manual

8

2SHUDWLQJ &RQWUROV DQG ,QGLFDWRUV

The ZOLL Base PowerCharger

4x4

with QuickCharge

illuminates the Batt. Ready indicator when the battery is
fully charged. The full charging cycle is complete in
four (4) hours or less for the PD 4410/Smart/Smart
Ready Battery and eight (8) hours or less for all XL
Batteries. The QuickCharge device provides no
information regarding battery capacity because it does
not test the battery capacity with each recharge. The
ZOLL Base PowerCharger

4x4

with QuickCharge relies

on the user to manually initiate the battery test cycle at
recommended intervals (see “Manual Battery Testing
Procedure” on page 9)
in order to confirm adequate
battery capacity.

 32:(5

Illuminates when the ZOLL Base PowerCharger

4x4

is

connected to live AC mains and the unit is ready for
charging and/or testing.

 &+$5*(5 21

Illuminates when a battery is properly installed in the
corresponding charging/testing compartment and the
charge cycle is in progress.
There are four CHARGER ON indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above it.

 %$77 5($'< 4XLFN&KDUJH

Illuminates at the end of a charge cycle. Indicates that
the battery has been charged to 100% of its present
capacity and is ready for use. The BATT. READY light
does NOT give any indication as to the available run time
left in the battery.
Illuminates at the end of a battery test or battery
compartment test and remains lit as long as battery
remains in compartment. Indicates the battery
compartment or battery has passed the capacity test and
is operating properly.
There are four BATT. READY indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 )$8/7

Illuminates, with the TESTING light, when a battery
capacity test fails. Also, illuminates when a battery
compartment or any other failure occurs.
There are four FAULT indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 7(67,1*

Illuminates when either a battery capacity test cycle is in
progress or a battery compartment test is completed.

Illuminates with the BATT. READY light indicating a
usable fully charged battery, capable of powering an
M Series equipped with Pace/Defibrillation and Pulse
Oximetry (SpO

2

) for approximately 1.5 hours in Monitor

mode.
Illuminates with the FAULT light to indicate that a battery
has failed the capacity test.
There are four TESTING indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 7(67 EXWWRQ

Initiates a battery test cycle or battery compartment test.
The battery capacity test cycle automatically charges,
discharges and recharges the battery.
If a battery is installed in the battery compartment and
the corresponding TEST button is pressed, a battery
capacity test is performed. If the battery compartment is
empty and the corresponding TEST button is pressed, a
compartment test is performed and an audible tone will
sound, indicating that the test has passed.
There are four TEST buttons. Each one corresponds to
the battery charging/testing compartment directly above.

6HWXS

Place the ZOLL Base PowerCharger

4x4

on a stable

secure surface. There must be at least four (4) inches of
clear space around the unit to allow for proper heat
dissipation during charging or testing batteries. Do not
place anything on top of or beneath the unit, such as
blankets or cloths. Doing so may block the vents on the
unit, preventing proper dissipation of heat during
operation.
Turn the unit on by plugging the line cord into the power
module at the rear of the unit and then into live AC
mains.

%DWWHU\ &KDUJLQJ 3URFHGXUH

4XLFN&KDUJH

The ZOLL PD 4410/Smart/Smart Ready Battery may be
charged in any of the four battery compartments.
XL Batteries should only be charged in a ZOLL Base
PowerCharger

4x4

labeled “XL Battery Ready”. Each

battery charge compartment operates independently of
the others. Four batteries may be charged
simultaneously. Installation or removal of adjacent
batteries will not affect the ongoing charging of batteries
in other compartments. Correct placement of the battery
into any compartment will automatically initiate the
charging cycle.

Advertising
This manual is related to the following products: