Manual battery testing procedure, 0dqxdo %dwwhu\ 7hvwlqj 3urfhgxuh, Operation (quickcharge) – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual

Page 13

Advertising
background image

9

Operation (QuickCharge)

To turn the ZOLL Base PowerCharger

4x4

on, plug the

power cord into live AC mains. All indicator lights will
illuminate for approximately five (5) seconds and then
go off. An audible tone will sound indicating self test
passed and the POWER light will remain lit.
To charge a battery:
1. Check to see that the unit is connected to live AC

mains. The POWER light should be lit.

2. Insert a battery into any compartment. Make sure the

battery is fully seated (a click can be heard).

3. The CHARGER ON indicator illuminates and remains

on during the charging cycle.

4. In four (4) hours or less for the PD4410/Smart/

Smart Ready Battery and eight (8) hours or less for
all XL Batteries, the CHARGER ON light will go off
and the BATT. READY light will illuminate indicating
the battery has been charged to 100% of its present
capacity and is ready for use.

If the battery or battery compartment is not capable of
charging, the FAULT light will illuminate and the
CHARGER ON light may stay on.

To identify the defect, press and hold the TEST
button while the battery is still in the compartment.

Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 11 and
take corrective action.
Before disposing of the battery, test the battery
compartment for defective operation (see “Battery
Compartment Testing Procedure” on page 10) and
perform a battery test.

1RWH Interrupting a charge cycle in progress by

removing the battery will cause an audible
warning of five (5) consecutive beeps to indicate
that the battery is not fully charged and should not
be used. In addition, all four indicator lights will
illuminate for four (4) seconds and then go off.

Repeated use of partially charged batteries will
reduce battery capacity and operating life.

5. To remove the battery, insert a finger into the recess

at the top of the battery compartment, press against
the battery to disengage the battery locking clip and
lift the battery out. A compartment test is performed
illuminating all indicator lights, briefly followed by the
BATT. READY and TESTING indicator lights
illuminating for two (2) seconds and then all indicator
lights going off, except the POWER indicator light.

0DQXDO %DWWHU\ 7HVWLQJ 3URFHGXUH

ZOLL batteries are sealed lead-acid batteries. It is
recommended that all batteries be tested for capacity at
least every 90 days and that the test results and date be
written on the battery label.
All four battery compartments can test batteries. It is not
necessary to insert a fully charged battery for testing; the
ZOLL Base PowerCharger

4x4

will automatically charge

the battery first, test it, and then recharge the battery to
full capacity. Testing a battery in one compartment will
not affect any other charge or testing cycles in progress.
To test a battery:

1. Check to see that the ZOLL Base PowerCharger

4x4

is connected to live AC mains. The POWER indicator
light should be lit.

2. Insert a battery into any compartment. Make sure the

battery is fully seated (a click can be heard).

3. Press the TEST button for the appropriate

compartment. The TESTING indicator light will

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON