Battery compartment testing procedure, Dwwhu\ &rpsduwphqw 7hvwlqj 3urfhgxuh – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual

Page 14

Advertising
background image

ZOLL Base PowerCharger

4x4

Operator’s Manual

10

illuminate for the duration of the test cycle.

4. If the battery passes the test, the BATT. READY light

will illuminate while the TESTING light remains lit.
The battery is now recharged, ready for use, and
capable of powering an M Series equipped with
Pace/Defibrillation and Pulse Oximetry (SpO

2

) for

approximately 1.5 hours in Monitor mode.

5. If a battery fails the test, the FAULT light will

illuminate while the TESTING light remains lit.
To identify the defect, press and hold the TEST
button while the battery is still in the compartment.
Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 11 and
take corrective action.
Repeat test.
Before disposing of the battery, test the battery
compartment for possible defective operation (see
“Battery Compartment Testing Procedure” on
page 10).

6. Record the date and results of the test on the battery

label.

1RWH Interrupting a charge cycle in progress by

removing the battery will cause an audible
warning of five (5) consecutive beeps to indicate
that the battery is not fully charged and should
not be used. All four indicator lights will illuminate
for four (4) seconds and then go off.

%DWWHU\ &RPSDUWPHQW 7HVWLQJ

3URFHGXUH

The ZOLL Base PowerCharger

4x4

has four battery

compartments that are able to test and charge batteries
simultaneously and independently of each other. In
addition, the ZOLL Base PowerCharger

4x4

can test each

battery compartment for defective operation through the
simple press of a button.
To test a battery compartment:
1. Check to see that the unit is connected to live AC

mains. The POWER light should be lit.

2. Remove the battery from the compartment to be

tested.

3. Press the TEST button under the battery

compartment.

4. All four indicator lights will illuminate for

approximately 4-5 seconds.

5. If the TESTING light and BATT. READY light remain

lit for three (3) seconds while the other lights go off,
the compartment is fully operational and has no
defects.

6. If the FAULT light is lit, the compartment is defective

in some way.
To identify the defect, press and hold the TEST
button. Illumination of certain indicator lights together
represent a specific defect. See “Fault Codes and
Definitions” on page 11
to identify the fault and take
corrective action.

1RWH Battery compartments operate independently of

each other and one faulty compartment may not
affect any others. However, there are common
failures that can affect all compartments.

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

Advertising