Taking Measurements
9650-1214-01 Rev. G
11
When reading the blood pressure values on the display, 
keep in mind that the following conditions can influence 
NIBP measurements: 
•
Position of the patient
•
Position of the cuff relative to the patient’s heart
•
Physical condition of the patient
•
Patient limb movements
•
Convulsions or tremors
•
Very low pulse volumes
•
Premature ventricular beats
•
Vibrations in the cuff or hose caused by moving 
vehicles
•
Improper cuff size or application
Taking a Single Measurement
You can take a single measurement by pressing the 
NIBP button unless the E Series defibrillator is charged 
or charging.
To take a single measurement, press the NIBP button on 
the E Series front panel. 
The E Series unit takes one blood pressure 
measurement.
Taking STAT Measurements
You can set the E Series unit to take a STAT 
measurement, which consists of the unit taking as many 
blood pressure measurements as possible (up to ten) in 
five minutes. When the E Series unit finishes taking a 
measurement and the cuff deflates, it immediately starts 
another measurement. Whenever practical, allow 
several minutes between STAT measurement 
sequences to restore full circulation to the monitored 
limb.
You cannot set the E Series unit to take STAT 
measurements while the defibrillator is charged or 
charging. Charging the defibrillator aborts all STAT 
measurements and deflates the cuff.
Starting STAT Measurements
To start STAT measurements, either:
• Press the NIBP button and hold for two seconds (if
the unit is so configured), or
• Press the NIBP Stat softkey.
If the NIBP Stat softkey label is not displayed, see 
“Displaying the NIBP Menu” on page 8.
If the E Series unit is configured to automatically 
generate strips, it prints a history of the STAT 
measurements at the end of the five-minute period.
Aborting STAT Measurements
You can abort STAT measurements at any time. Doing 
so terminates the current measurement and all 
subsequent measurements.
To abort STAT measurements, either:
• Press the NIBP button, or
• Press the NIBP Stat softkey.
If the NIBP Stat softkey label is not displayed, see 
“Displaying the NIBP Menu” on page 8.
The E Series unit immediately aborts all measurements, 
and the cuff deflates.
Taking Automatic Measurements
You can set the E Series unit to automatically take a 
series of measurements at selected intervals. For 
example, if you set the E Series unit to take an automatic 
measurement with an interval of 15 minutes, it 
immediately takes a measurement, waits 15 minutes, 
WARNING
If the E Series unit takes a measurement but detects 
the presence of artifact in the signal (denoted by “*” in 
the NIBP display area), the measurement may not be 
accurate. Under such circumstances, perform 
additional blood pressure measurements. If you 
repeatedly obtain artifact, use alternate techniques to 
determine blood pressure prior to taking clinical 
action.
CAUTION
Make sure that no measurement is being taken when 
you start a new measurement. If you press the NIBP 
button while the E Series unit is taking a 
measurement, the measurement aborts immediately 
and the cuff deflates.
#(!2'%2
35--!29
#/$%
-!2+%2
.)"0
021,725
(&*
1,%3
6S2
&2
P
P
+J
55
(&*
,,
[
,%!$
3):%
(2
3DUDP
:DYH
$ODUPV
/HDG
,'
NIBP Button
WARNING
Repeated use of STAT measurements on the same 
patient over a short time interval can affect blood 
pressure readings, limit circulation to the limb, and 
cause injury to the patient.