ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual
Page 16

E S
ERIES
- N
ONINVASIVE
B
LOOD
P
RESSURE
(NIBP) O
PTION
I
NSERT
16
9650-1214-01 Rev. G
REPEAT NIBP MEAS
alternating with
NIBP SIGNAL WEAK
There is a weak or no oscillometric signal.
Check cuff fit and positioning.
Check hose connection to unit.
Check for kinked hose.
Increase cuff inflation pressure if clini-
cally appropriate.
SYSTEM FAULT
Communication problem with the NIBP mod-
ule.
Cycle power, and retry.
If problem persists, return for service.
False high reading
Cuff too small.
Cuff not centered over brachial artery.
Cuff too loose.
Patient’s limb below level of heart.
Select larger cuff.
Readjust cuff.
Reapply cuff or select smaller cuff.
Raise patient’s limb to heart level.
False low reading
Cuff too large.
Limb above level of heart.
Select smaller cuff.
Lower patient’s limb to heart level.
Message/Symptom
Problem
User Action