3. Adjust the cuff so that the artery marker on the cuff is
over the artery, pointing to the hand or foot.
4. Check that the cuff ends between the range lines
marked on the cuff.
If they do not line up, use a different size cuff.
5. Wrap the deflated cuff snugly around the limb without
impeding blood flow.
6. Ensure that the hose is routed to avoid kinking or
compression.
You can now access the NIBP features.
Accessing NIBP Features
Unless you are sure that the NIBP cuff inflation and 
alarm settings are appropriate for the patient, display the 
NIBP menu before you take a blood pressure 
measurement. When you first turn on the E Series unit, 
the NIBP settings are at their default values. 
While the factory-installed default settings are 
appropriate for most patients, do not assume the settings 
are at their default. A previous user may have:
•
Changed the settings (if you did not turn on the 
E Series unit).
•
Reconfigured the default settings.
Use the default settings unless they are clearly 
inappropriate for the patient. Any changes to these 
settings remain in effect until either the settings are again 
changed, or for 10 seconds after the E Series unit is 
turned off. If you have not received training on setting 
NIBP features, do not use the NIBP option. 
To facilitate quick reaction during emergency situations, 
you can directly access many NIBP features without 
displaying the NIBP menu (see the following table).
Displaying the NIBP Menu
Unless it is an emergency situation where quick reaction 
is essential, you should always check that the cuff 
inflation and alarm settings are set properly before taking 
a measurement. You can access all NIBP features 
through the NIBP menu. You can display the NIBP menu 
without leaving MONITOR, DEFIB or PACER mode. 
AEDs must be in manual mode to display the NIBP 
menu.
To display the NIBP menu:
1. Press the Param softkey.
If the Param softkey label is not displayed, press the 
Return softkey until it is displayed.
The E Series unit displays the Parameter menu:
2. Press the Select softkey until NIBP is highlighted.
3. Press the Enter softkey.
The E Series unit displays the NIBP menu:
CAUTION
Do not place the NIBP cuff on the same arm/leg as an 
SpO
2
sensor. Inflation of the cuff causes the SpO
2
monitor to read incorrectly. Also, do not attach the cuff 
to a limb being used for IV infusion. Cuff inflation 
might block the infusion, causing harm to the patient.
CAUTION
Ensure that the cuff is at the same level as the heart. 
If the cuff is located below the patient’s heart, the 
blood pressure measurements display false high 
readings. If the cuff is located above the patient’s 
heart, the blood pressure measurements display false 
low readings. 
WARNING
Do not start NIBP measurements unless you are sure 
that the cuff inflation and alarm settings are 
appropriate for the patient. Incorrect settings can 
result in patient injury or inaccurate measurements.
Task
Action
Taking a single 
measurement
Press the NIBP button.
Taking STAT
measurements
Press and hold the NIBP
button for two seconds (if the 
NIBP Button Press and Hold 
configuration option is set to 
NIBP STAT), or see the next 
section, “Displaying the NIBP 
Menu.”
Taking automatic 
measurements
Press and hold the NIBP
button for two seconds (if the 
NIBP Button Press and Hold 
configuration option is set to 
NIBP AUTO), or see the next 
section, “Displaying the NIBP 
Menu.”
Aborting measure-
ment in progress
Press the NIBP button.
Changing NIBP
settings
See “Displaying the NIBP 
Menu” on page 8.