Nibp indications for use, Contraindication for use, What can i do with the nibp option – ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual
Page 5: How do i use the nibp option, How does the nibp option work
NIBP Indications for Use
9650-1214-01 Rev. G
5
NIBP Indications for Use
The ZOLL E Series
®
NIBP option is indicated for the
non-invasive measurement of arterial blood pressure for 
resting patients in critical care and transport. The NIBP 
option is designed to measure blood pressure for adult 
and pediatric patients.
Contraindication for Use
The ZOLL E Series NIBP option is not indicated for use 
on neonatal patients or infants whose upper arm 
circumference is less than 13 cm.
What Can I Do With the NIBP Option?
With the NIBP option you can do the following in 
MONITOR, PACER, or DEFIB mode:
•
Take a single blood pressure measurement.
•
Take a STAT measurement (as many measurements 
– up to ten – as possible within five minutes).
•
Take repeated measurements at user-programmable 
intervals.
•
Immediately abort any measurement in progress by 
pressing one button.
•
Set the cuff inflation pressure to adjust automatically 
based on the previously measured systolic value.
•
Display systolic, diastolic and mean blood pressure 
on the screen.
•
Configure alarms to go off when the unit detects 
blood pressure values above or below 
user-programmable limits.
How Do I Use the NIBP Option?
To take safe and accurate blood pressure measurements 
using the E Series NIBP option, you must perform the 
following steps, each of which corresponds to a 
subsequent section in this document. Read each section 
carefully before you use the E Series NIBP option.
1. Select the proper size cuff.
2. Connect the hose to the E Series unit and to the cuff.
3. Apply the cuff to the patient.
4. Display the NIBP menu.
5. Set the cuff inflation pressure (if the current setting is
not appropriate).
6. Configure alarms (if the current settings are not
appropriate).
7. Take blood pressure measurement(s).
8. Read the display.
How Does the NIBP Option Work?
The patient blood pressure cuff and hose connect to the 
E Series unit through the NIBP connector on the rear 
panel of the unit. The NIBP button on the front panel of 
the E Series allows you to initiate and terminate blood 
pressure measurements, whose values display in the 
NIBP area of the monitor. You can also initiate and 
terminate measurements using the softkeys on the NIBP 
menu.
The NIBP module measures the oscillometric pulses 
transmitted through the blood pressure cuff and hose, 
and calculates the blood pressure measurements 
accordingly. The pressure measurement cycle takes 
approximately 30 seconds and proceeds as follows:
1. The cuff inflates to a preconfigured pressure
(configured default value for adult patients is 
180 mmHg), above the patient’s systolic blood 
pressure, to occlude blood flow through the arteries 
in the monitored limb.
2. The cuff deflates incrementally, allowing blood to flow
through the cuff and into the monitored limb.
3. As blood flows through the partially deflated cuff, it
produces pressure oscillations that are transmitted to 
the E Series unit through the hose.
4. The E Series unit measures the oscillometric pulses
and uses them to calculate the corresponding 
systolic, diastolic, and mean blood pressure, as well 
as the patient’s pulse rate.
5. The NIBP option automatically adjusts the blood
pressure measurement procedure in response to 
certain error conditions such as:
WARNING
Do not use the NIBP option without proper training. 
Setting initial inflation pressure too high can result in 
serious injury to the patient. Patient movement, very 
low pulse volume, or vibration from outside sources 
can influence the accuracy of blood pressure mea-
surements.