The reduction in fever burden was accompanied by a reduction in the use of
adjunctive cooling means as presented in the tables below.

Table 4. Use of Topical Cooling Devices

Cool Line

Control

% Reduction

p*

One or more topical cooling
device (n/N, %)

26 / 154

16.9

67 / 142

47.2

64%

<0.0001

Cooling Blanket use (n/N, %)

25 / 154

16.2

59 / 142

41.6

61%

<0.0001

Other device use (n/N, %)

7 / 154

4.6

19 / 142

13.4

66%

0.008

* Fisher’s exact test