Table 9. catheter characteristics, Table 10 – ZOLL Thermogard XP IVTM Physician Manual User Manual

Page 25

Advertising
background image

ZOLL IVTM™ System

Physicians' Manual

600248-001 Rev 3

24

Table 9. Catheter Characteristics

Catheter

Effective Length

Infusion Lumen

Nominal Power

(cooling / heating)

Icy

38 cm

Proximal, Middle and

Guidewire (Distal) lumen

140 / 35 W

Quattro

45 cm

Proximal, Middle and

Guidewire (Distal) lumen

180 / 50 W

Solex

25 cm

Proximal, Middle and

Guidewire (Distal) lumen

140 / 35 W

Anesthetized patients are unable to turn on normal heat generation (e.g. through
shivering or adrenergic drive) and have impaired peripheral vascular responses i.e.
they are rendered essentially poikilothermic by the anesthetic. Under these
conditions, these catheters will provide approximately the following rates of cooling
and heating in a 75 kg person:

Table 10.

Catheter

Cooling

Heating

Icy

2 ºC / hr

0.5 ºC / hr

Quattro

2.5 ºC / hr

0.8 ºC / hr

Solex

2 ºC / hr

0.5 ºC / hr

The Icy and Quattro catheters are femoral lines.

The Icy has a lower surface area and is therefore theoretically less likely to cause
thrombo-embolic events. It is also significantly shorter in its applicable length.