Lead summary reports, Lead ecg waveforms – ZOLL M Series Defibrillator Rev B Non-Int 12 Lead User Manual

Page 10

Advertising
background image

OPTION INSERT

9650-0218-01 Non-Interpretive

12-lead

-

8

12-Lead Summary Reports

Immediately after acquisition, the 12-lead data prints in the following order:

1. 12-Lead

ECG

waveforms

2. Patient

information

See the figures below for a view of each print-out.

12-Lead ECG Waveforms

4x3 Format

This format prints 10 seconds of 12-lead ECG data, in four staggered 2.5 second segments. Each 2.5 second segment displays
the ECG reading for one set of three leads (see figure). The report can be configured for either standard or Cabrera print formats.
1 mV calibration pulses are printed at the beginning of the report for each data channel. The M Series can be configured to print
2.5, 5, 7.5, or 10 seconds of ECG data for each lead. The default is 2.5 seconds per lead. See the M Series Configuration
Guide, P/N 9650-0201-01
for instructions on changing this default setting.

Standard 12-lead printout is always 4x3 Format unless configured differently.

1

st

2.5 sec ECG data 2

nd

2.5 sec ECG data 3

rd

2.5 sec ECG data 4

th

2.5 sec ECG data