ZOLL M Series Defibrillator Rev B Non-Int 12 Lead User Manual

Page 8

Advertising
background image

OPTION INSERT

9650-0218-01 Non-Interpretive

12-lead

-

6

0.05– 40 4x3 channel printout (0.05 – 40Hz 4x3): When
programmed to this setting, all 4x3 12-lead ECG reports will
print with a 0.05 to 40 Hz bandwidth. This setting can be used
to reduce excessive muscle or other artifact.

Select the filter to be used by pressing the Filter softkey. The
highlight area will scroll among the different filter choices.
Press the Enter softkey to save the highlighted filter and
return to the 12 Lead Monitor submenu.

NOTE: When the M Series is turned off for more than 10
seconds, all settings are restored to their original default
settings.

Lead Group Setting

The M Series non-interpretive 12-Lead Option allows any
three ECG lead signals to be simultaneously printed on the
strip chart recorder when using a 12-lead cable. When the
Standard lead grouping is selected, leads I, II and III are
printed together, leads aVR, aVL, and aVF are printed
together, leads V1, V2, and V3 are printed together and leads
V4, V5, and V6 are printed together. When either Custom 1 or
Custom 2 lead grouping is selected, the three (3) leads
printed on the recorder will be those previously programmed
for the custom group via the Configuration mode of the M
Series device. (See the M Series Configuration Guide, P/N
9650-0201-01,
for more information.)

Standard

When programmed to this setting, 3 -lead ECG recordings will
be grouped as follows:

• leads I, II, and III
• leads aVR, aVL, and aVF
• leads V1, V2, and V3
• leads V4, V5, and V6.

The lead group that prints on the recorder in this mode will be
the group containing the lead selected for display on the
M Series screen. e.g. If Lead II is displayed, leads I, II and III
will be printed. If lead V3 is di splayed, leads V1, V2, and V3
will be printed.

NOTE: When the M Series is turned off for more than 10
seconds, all settings are restored to their original default
settings.

Custom

When the lead group is configured to either Custom 1 or
Custom 2, 3-lead ECG recordings will include the leads
preconfigured for that custom group.

Pressing the Lead Grp softkey will scroll the highlight among
the different selections allowing the user to select the
appropriate Lead Group for three channel printing.

Press the Enter softkey to save the highlighted Lead Group
and return to the 12 Lead Monitor menu.

Dial Type Setting

Selecting the Dial Type setting allows the user to select either
Tone or Pulse dialing.

Pressing the Dial Typ softkey toggles the highlight between
both types of dialing modes. Press the Enter softkey to save
the highlighted dial mode and return to the 12 Lead Monitor
menu. Press the Return softkey to return to the 12 Lead
Monitor menu without saving the highlighted dial mode.

PT Info Softkey

Press the PT INFO softkey to access the patient demographic
information.

Pressing the PT INFO softkey allows the user to select from
the following softkeys: ID#, Age, Gender, Patient Records or
Return.
Note: If no information has been entered, the default age of
“60” will be displayed above the Age softkey and the default
gender of “Male” will be displayed above the Gender softkey.

Advertising