Warnings – ZOLL M Series Defibrillator Rev B Non-Int 12 Lead User Manual

Page 4

Advertising
background image

OPTION INSERT

9650-0218-01 Non-Interpretive

12-lead

-

2

WARNINGS

• Before use, carefully read the M Series Operator’s Guide and these operating instructions.
• The M Series with non-interpretive 12-lead ECG is designed to acquire ECG data from resting, supine patients. Use of

the device to acquire ECG signals from moving or shaking patients may produce erroneous 12-lead data. Always

ensure that the patient is kept motionless during 12-lead ECG signal acquisition.

• Excessive body hair or wet, sweaty skin may interfere with electrode adhesion. Remove the hair and/or moisture from

the area where the electrode is to be attached.

• Using previously opened or out of date electrodes may degrade the ECG signal quality. ECG electrodes should be

removed from their sealed package immediately prior to use.

• Monitoring electrodes may become polarized during defibrillator discharge, causing the ECG waveform to briefly go off

screen. ZOLL Medical Corporation recommends the use of high quality silver/ silver chloride (Ag/AgCl) electrodes to

minimize this effect, and circuitry in the instrument will return the trace to the monitor display within a few seconds.

• Wait 15 seconds after defibrillator discharge before attempting a 12-lead acquisition. Electrode polarization subsequent

to defibrillator discharge may result in excessive noise on the 12-lead ECG printout.

• Cover the patient cable's V-lead connector with the plastic cap supplied when the V-leads are not in use. Failure to do

so may result in a shock hazard during defibrillation attempts.

• To assure protection against the effects of defibrillator discharge use only 12-lead cables supplied by ZOLL Medical

Corporation.

• To avoid a shock hazard and interference from nearby electrical equipment keep electrodes and patient cables away

from grounded metal and other electrical equipment.

• Do NOT sterilize the M Series unit or any of its accessories except internal handles or as directed.
• Check the operation and integrity of your M Series and 12-lead cable regularly by performing the Daily Operational

Verification Test.

• Grounding reliability can ONLY be achieved when the equipment is connected to an equivalent receptacle marked

“HOSPITAL ONLY” or “HOSPITAL GRADE”.

• The frequency response of the monitor screen is intended only for basic ECG rhythm identification; it does not provide

the resolution required for diagnostic and ST segment interpretation. Use the strip chart recorder for th is purpose.

• Implanted pacemakers may cause the heart rate meter to count the pacemaker rate during incidents of cardiac arrest or

other arrhythmias. Pacemaker patients should be carefully observed. Check the patient's pulse; do not rely solely on

heart rate meters. Dedicated pacemaker detection circuitry may not detect all implanted pacemaker spikes. Patient

history and physical exam are important in determining the presence of an implanted pacemaker.