Non-Interpretive 12-lead

9650-0218-01 Non-Interpretive

12-lead

-

7

Patient Identification Number (ID#)

Pressing the ID # softkey causes the “Name and ID#” screen
to appear. “Name” is entered by pressing the Prev Digit and
Next Digit softkeys to select the digit then pressing the Inc
Digit
and Dec Digit softkeys to select the value of the digit.
When “Name” has been entered, press the Enter Name
softkey. The highlight will automatically advance to the “Patient
ID#” line. The Patient ID# is entered exactly the same way as
above. When the ID# has been entered, press the Enter ID
and Return
softkey to return to the Patient Information Menu
and store the name and ID number. If no patient ID # is
entered the M Series automatically generates a 12 digit patient
ID# based on the year/date/time (2000021513 20) of the first
12-lead acquisition.

Patient Age

Pressing the Age softkey causes the number displayed above
the Age softkey to be highlighted. Press the Dec Age or Inc
Age
softkey to change the patient’s age. Press the Enter
softkey to save the highlighted age and return to the Patient
Information screen. Press the Return softkey to return to the
Patient Information screen without saving the highlighted age.

Patient Gender

Pressing the Gender softkey toggles the highlight between
both genders displayed above the Gender Softkey. Press the
Enter softkey to save the highlighted gender and return to the
Patient Information screen. Press the Return softkey to return
to the Patient Information screen without saving the
highlighted gender.

Patient Records

Pressing the Patient Records softkey causes the Patient
Records screen to appear. This screen is used to transmit or
print specific 12-lead ECG records denoted by Patient ID #,
date, and time. (Patient Records are stored in Summary
memory. The user may wish to set Summary

Report Restart

Delay to 1.5 days in order to extend the storage time for 12 -
lead Patient Records. See the M Series Configuration
Guide, P/N 9650-0201-01,
for instructions on how to
implement this feature.)

Pressing the Prev Record and Next Record softkeys will
scroll the highlight up and down through the list of stored 12 -
lead records.

Pressing the Print softkey will print the selected 12 -lead
patient record on the strip chart. The format is the s ame as
that stored in summary memory.

Pressing the Transmit softkey will transmit the selected 12-
lead patient record. The Transmission Setup screen will
appear as described in the Data Transmission section of this
manual.

Pressing the Return softkey exits the Patient Records screen
and displays the 12 Lead menu.