9650-000840-01 Rev. A

SurePower

TM

ll Battery Pack

Guide

1

Introduction

The SurePower

TM

II Battery Pack is a rechargable, sealed lithium ion battery for use in

compatible ZOLL

®

defibrillator/monitors and chargers.

It is important that you read and understand the information in this manual. The proper use of
the SurePower II Battery Pack is critical to the reliable operation of your ZOLL defibrillator/
monitor.

Conventions

This manual uses the following conventions:

Warning statements describe conditions or actions that can result in personal injury or
death.

Caution

Caution statements describe conditions or actions that can result in damage to the SurePower II

Battery Pack.

Note: Notes contain additional information on using and maintaining the SurePower II

Battery Pack.

The manual displays button and LED labels in bold face type.

Safety

The following sections provide important warnings, cautions, and notes regarding the ZOLL
SurePower II Battery Pack.

Warnings

Locate ZOLL defibrillator/monitor units where they cannot harm the patient or medical

personnel if they fall -- accidental sliding or drop of the SurePower II Battery Pack may

cause physical injury. Damaged batteries may expose users to a safety hazard. Should the

SurePower II Battery Pack be dropped, inspect and test the unit prior to further use

following the procedures described in the Maintenance section of this Guide.

Avoid prolonged exposure of the SurePower II Battery Pack to direct sunlight to minimize

heating and damage that may occur from exposure to ultraviolet light.

SurePower II Battery Packs have been tested against interference from radio frequency

emissions typical of two-way radios used in emergency service/public safety activities.

Refer to the appropriate ZOLL defibrillator/monitor operator’s guide to determine

recommended operating distances from RF transmission equipment

.

Do not use SurePower II Battery Packs unless they have been visually inspected, tested, and

recalibrated by appropriate personnel and you have been adequately trained in the proper

use of these batteries. See the Maintenance section of this guide for more information.

Do not disassemble the SurePower II Battery Pack.

Charge the SurePower II Battery Pack in a well-ventilated area.

You must perform periodic inspection, capacity testing, and functional testing of

rechargable defibrillator batteries to ensure the availability of safe and reliable batteries. See

the Maintenance section of this guide for more information.