Yokogawa 202 Field Mounting Rate Totalizer User Manual
Page 14

Step
Comment
CAL 06
Filter.
Fluctuations in the flowrate can be filtered out so that
the Rate is held steady. The filter value is programmed
between 1 - 99 where "1" represents no filtering and 99
is maximum.
See section 3.4.
CAL 07
Decimal Point for Total Display.
The totals can be displayed with 0, 1, 2 or 3 decimal
points.
CAL 08
Total Conversion Factor - whole numbers.
CAL 09
Total Conversion Factor - digits after the decimal point.
The Total Conversion Factor enables the Rate to be
displayed in one engineering unit and the Totals to be
displayed in another engineering unit. The Total
Conversion Factor can be programmed in the range of
0.000001 - 999,999.
Set to 1.000 if totals and rate are in the same unit eg.
litres.
See section 3.3.
CAL 10
Low Alarm - whole numbers.
12 Programming