1 programming the setup parameters – Yokogawa 414 Batch Controller User Manual

Page 34

Advertising
background image

32 Calibration

5.1 PROGRAMMING THE SETUP PARAMETERS

Step

Display

Description

Text

Ref

1

CAL

BATCH
OPTION
TEST
END

Select the Calibrate mode to setup program
parameters.
Select Batch to enter Batch Setup parameters.
Option (if installed).
Select the test mode to check input signals.
Exit to normal operation.

5.2
5.3
5.4

The following steps are displayed if CAL is selected.

2

RESTOT

Reset all totals to zero.
To reset all totals (resettable and accumulated)
press the BATCH SET key once.

3

SCALE

Scaling Factor.

Fact

Enter the Scaling factor (K-factor) of the
flowmeter.

3.3.1

4

F dPt

Number of decimal points with which the Rate
is to be displayed between 0 to 0.00000.

5

t.base

The Timebase with which the Rate is
calculated must be entered as:

60secs
hours
days
secs

units/min
units/hour
units/day
units/second

6

FILTER

The filter constant for filtering the rate display. 3.3.2

1

No filtering.

to
99

Very heavy filtering.