<4. GETTING STARTED>

4-4

IM 11M12D01-61E

Table 4.2 Parameters required for operation

Figure 4.2 Available Address Range

4.3 Turning on Power to AV550G and Fieldbus

Turn on power to the host, bus and AV550G. If the display of the AV550G does not light up, or if an

abnormal current flows from the power supply, check the power supply voltage.

Use the host’s display to confirm that the AV550G is operating normally on the bus. Unless advised

otherwise, you can assume that the AV550G factory default Fieldbus device settings are as follows:
PD tag: AV550G

Node address: 247 (hexadecimal F7)

Device ID: 5945430401xxxxxxxx (xxxxxx = six alphanumeric characters)

T0401.ai

Symbol

Parameter name

Description and Value

Set a value of 4 or greater

Set a value of 4 or greater

Slot-Time

V (ST)

First-Unpolled-Node

Minimum-Inter-PDU-

Delay
Maximum-Reply-

Delay

This sets the number of

consecutive unpolled

nodes.

Factory default setting is

F7 (hex).

This address determines

the number of addresses

Reserved for BASIC

devices, as shown in Fig.

4.2

Defines the first address

that can be used by host.

Set a value of 15 or

greater (hex).

Number-of-

consecutive-

Unpolled-Node

Set this such that V(MRD)

x V(ST) is 12 or greater.

V (MRD)

V (MID)

V (FUN)

V (NUN)

Bridge devices

Do not use

0x10

0x00

0x0F

0x14

0x13

0xF7

0xF8

0xFB

0xFC

0xFF

V(FUN)

V(FUN)+V(NUN)

LM devices

Unused

V(NUN)

Note 1: LM device: with bus control function (Link Master function)

Note 2: BASIC device: without bus control function

BASIC devices

Default addresss

Portable device address

F0403.ai

AV550G (0xF7)