Preparations before Starting Measurement (Set up the PR300 First)

2-3

IM 77C01E01-01E

1

2

3

4

5

A

I

2.2

Setting the Phase and Wire System

Measured Value screen

Startup screen

The PR300 shows the Startup
screen for about 5 seconds, then the
Measured Value screen appears.

Setting completed.

NOTE

If you change the phase and wire system, all parameters other than those related to RS-485 and
Ethernet communications are initialized (to factory-set values). Change the phase and wire system
before setting parameters such as the VT and CT ratios.

Range of Phase and Wire System Options

Parameter

Symbol

Model and Suffix Codes

Parameter

Name

Setting Type

Setting Range (Details)

Initial Value

(Factory-set

Value)

Phase and

wire system

Selection

Single-phase
two-wire system

Single-phase
three-wire system

Three-phase
three-wire system

Single-phase
two-wire system

Single-phase
three-wire system

Three-phase
three-wire system

Three-phase
four-wire system

Three-phase
four-wire system
(2.5 element)

Three-phase

three-wire

system

Three-phase

four-wire

system

Three-phase

four-wire system

(2.5 element)

NOTE

If single-phase three-wire system is selected, the voltage range is fixed at 300 V (between P0 and

P1, P0 and P2). The voltage range cannot be selected.

Three-phase four-wire system (2.5 element) can be used only when the voltage is in a state of

equilibrium. In addition, the phase and wire system cannot be changed.