IGA 320/23-LO Operating Manual

Controls and Connections 15

3.2.3 Connecting additional analyzing devices

Additional analyzing instruments, such as a LED digital display instrument, need to be connected
to a power supply and the analog outputs from the pyrometer. Other instruments, like a
controller or printer, can be connected to the display in series as shown below (total load of
resistance max. 500 Ohm).

3.3 Mechanical Installation

3.3.1 Fiber optic

The fiber has either a blue mark

(MB 7) or a red mark (MB 12) for
correct connection to the pyrometer.
This color mark has to be mounted on
the pyrometer’s side.
Please note that the connector on the

instrument side of the fiber has an
alignment pin. When connecting to
the socket, you have to turn the
connector until the pin is in the right
orientation before you can fix the
connector with the knurled nut.

Fiber

Screw connection

Pin

Cut-out

Blue or Red mark

Attention: The light guide end of the fiber optic cable as well as the

socket/connector and the optical head must always be protected with the caps when
not connected!

Ambient temperature

The fiber and optical head can withstand ambient temperatures up to 200 °C without cooling on
the side of the optical head.

Attention: The temperature of the fiber and optical head must be at least 30 °C

lower than the measuring temperature to get a correct temperature reading.

Serial number

The original fiber has a serial number which is also on the pyrometer’s housing. If required, the
fiber can be replaced by a new original IMPAC-brand fiber. In this case, the pyrometer has to be
calibrated together with the fiber (service).

yellow

LED digital display

Controller

Power supply

green

brown

white

230V ~

24 V DC

°C

Writer