7 Data format UPP (Universal Pyrometer

Protocol)

Via interface and suitable communication software or via Test function of the InfraWin
software (see

Section 5.5

Basic settings

Test) commands can be exchanged directly with

the pyrometer.
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).
The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with ok + CR.
Every command starts with the 2-digit device address AA (e.g. "00"). This is followed by 2 small
command letters (e.g. "em" for level of emissivity ), finished with CR.
This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter
"X" is omitted, then the device resets with the current parameter.
A ? after the small command letters answers with the respective settings (only at setting
commands, not at enquiry commands).

Example:

Entry:

“00em“

+ CR

The

emissivity

setting

( ) of the device with the address 00 is returned.

Answer:

“0970“

+ CR means Emissivity = 0.97 or 97.0%

Description

Command

Parameters

Reading
temperature
value:

AAms

Output: XXXXX (dec., in 1/10 °C or °F)

(77770 = Instrument’s temperature too high
88880 = temperature overflow)

Reading

temperature
value repeated:

AAmsXXX

XXX = 000...999 (XXX = number of measuring values)

Emissivity:

AAemXXXX

XXXX = (0100 ... 1000‰) (decimal)

Transmittance:

AAetXXXX

XXXX = (0100 ... 1000‰) (decimal)

Ambient
temperature
compensation:

AAutXXXX

XXXX = value of ambient temperature, 4-digit, hex

XXXX e.g. FFEC corresponds to -20 degrees
- 99

dez

= FF9D

hex

means: automatic, no manual

compensation

Exposure time t

90

: AAezX

X = 0 ... 6 (decimal)

0 = intrinsic time constant of the device
1 = 0.01 s

3 = 0.25 s

5 =

3.00 s

2 = 0.05 s

4 = 1.00 s

6 = 10.00 s

Clear times of
the maximum

value storage:

AAlzX

X = 0 ... 9 (decimal)

0 = Maximum value storage off

1 = 0.01 s

4 =

1.00 s 7 = external deletion

2 = 0.05 s

5 =

5.00 s 8 = automatic deletion

3 = 0.25 s

6 = 25.00 s