8 Technical data

Temperature ranges:

100 ... 700 °C (MB 7)
150 … 1200 °C (MB 12)

Sub range:

Any range adjustable within the temperature range

minimum span 51 °C

Spectral range:

2 to 2.6 µm (main wavelength 2.3 µm)

IR detector:

Extended InGaAs

Optics:

Optical head Type I or Type II (short distances only, see table below)
fiber 0.4 mm (blue fiber mark)
fiber 0.2 mm (red fiber mark)

DIN connector electronics side SMA connector optics side

Power supply:

24 V DC (10 to 30 V DC), ripple must be less than 0.5 V

Power consumption:

Max. 1 W

Analog output:

0 to 20 mA or 4 to 20 mA (linear), switchable

Load:

0 to 500

Switch contact:

Opto relays; max. 50 V DC, 0.2 A; max. P

max

= 300 mW

Hysteresis:

2 to 20 °C

Digital Interface:

RS485 addressable (half duplex), baud rate 1200 up to 38400 Bd

Resolution:

0.1 °C on interface;
< 0.025% of the adjusted temperature sub range at the analog output

Isolation:

Power supply, analog output, and digital interface are galvanically isolated
from each other

Parameters:

Adjustable via interface:
Emissivity , transmittance , exposure time t

90

, clear times for max. / min.

max./min. value storage, analog output, sub temperature range, ambient
temperature compensation, pyrometer address, switch contact, hysteresis,
baud rate, wait time t

W

, targeting light

Emissivity :

10.0 to 100.0% adjustable via interface in steps of 0.1%

Transmittance :

10.0 to 100.0% adjustable via interface in steps of 0.1%

Exposure time t

90

:

2 ms (with dynamical adaptation at low signal levels);

adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum / minimum
value storage:

Built-in single or double storage. Clearing with adjusted time t

clear

(off; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), via interface or automatically with the next

measuring object

Uncertainty:*

(with =1, t

90

=1 s, T

amb.

=23°C)

Up to 400 °C: 2 °C

Above 400 °C: 0.3% of measured value in °C + 1 °C

(Note: the pyrometer must be operate at least 30 min before these values are valid)

Repeatability:*

( =1, t

90

=1 s, T

amb.

=23°C)

0.1% of measured value in °C + 1 °C

*

The temperature of fiber and optical head must be at least 30 °C lower than the measuring temperature to get a

correct temperature reading