4 settings / parameter descriptions, 1 factory settings, 2 basic range – LumaSense Technologies IGA 320/23-LO User Manual

Page 19: 3 material, 4 temperature display in °c or °f

Advertising
background image

IGA 320/23-LO Operating Manual

Settings / parameter descriptions 19

4 Settings / parameter descriptions

The pyrometer is equipped with a wide range of settings for optimal adaptation to the required

measuring condition and to measure the temperature correctly.

All settings can be read and set only in the pyrometer parameters window of the

software InfraWin. Adjusting the settings at the instrument is not possible (detailed
description of the software see Chapter

5

, Software InfraWin).

Selecting the pyrometer parameters window shows the current settings of the

pyrometer. Changing a value can either be done by typing a value in an input box or by
selecting a preset value from the list field.

4.1 Factory settings

Temperature display (°C / °F) = °C
Emissivity ( ) = 100%

Transmittance ( ) = 100%
Exposure time (t90) = min
Clear time (tClear) = off
max / min = maximum value storage preset

Analog output (out) = 0 ... 20 mA
LED targeting light = active at Power-on
Sub range same as basic temperature range
Ambient temperature compensation = auto
Address = 00
Limit switch = Off
Baud rate = 19200 Bd

4.2 Basic range

Basic range automatically displays the total range of the pyrometer and can not be changed.

4.3 Material

The Material setting allows you to store the names of different measuring objects with their

emissivity values and to recall them from the list. You can also add and change the materials
listing through the text file "mat.txt" (provided in the folder InfraWin in the standard

Documents and Settings path of Windows, normally C:\ Documents and Settings \ <user
name> \ InfraWin).

4.4 Temperature display in °C or °F

Choose whether the temperature should be displayed in °C (Celsius) or °F (Fahrenheit).

Advertising