IS 6 Advanced / IGA 6 Advanced Manual

Settings / parameter descriptions · 26

Note: Only via own communication program with interface command (not

possible with InfraWin, because InfraWin automatically detects a connected
pyrometer): If parameters should be changed simultaneously on all pyrometers, the
global Address 98 can be used. This allows you to program all pyrometers at the

same time, regardless of the addresses that have already been assigned. If the address
of a pyrometer is unknown, it is possible to communicate with it using the global
Address 99 (connect only one pyrometer).

4.9 Focused Distance

The focused distance value can be
adjusted at all distances between 210 mm
and 5000 mm using the Focus Adjustment
Screw on the back of the device.
The LED Distance Indicator Light will turn
red and the focused measuring distance in

mm will automatically be shown on the
Digital Display within a few seconds of
making an adjustment using the Focus
Adjustment Set Screw.
The focused distance can be viewed at
any time through the InfraWin software or by using the UPP Data Format commands.

4.10 Baud Rate

The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the

cable. A maximum cable length for 19200 Bd with RS485 is 2 km. The baud rate is reduced by
50% if the transmission distance is doubled.
Baud Rates can be set at 1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200.

4.11 Pyrometer Internal Temperature

The internal temperature of the pyrometer can be read through the PC interface using the
InfraWin software or by using the UPP Data format commands. It is a few degrees higher than
the ambient temperature due to the heat generated by the electronics.

When using the Laser Targeting Light, the targeting light is switched
off automatically if the internal temperature of the pyrometer exceeds
61 °C. This safety feature is used to prevent damage to the laser. It can
only be used again once the temperature falls below 61 °C.