IS 6 Advanced / IGA 6 Advanced Manual

Data format UPP

Ò

· 36

Description

Command

Parameters

Response Time t

90

AAezX

X=0 to 9

0=min

1=1 ms 2=3 ms

3=5 ms

4=10 ms

5=50 ms

6=250 ms

7=1 s

8=3 s

9=10 s

Temp Display °C or °F AAfhX

Output: X=0 Display

in

°C

X=1

Display in °F

Device address

AAgaXX

XX=(00 to 97)
00 to 97=regular device addresses

99=global address with response

Internal temperature
(read)

AAgt

AAtm

Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F)
gt=current temp.

tm=maximum temp. (memory)

Laser

AAlaX
AAla

X=0 off X=1 on
Answer: 1-digit: “0” or “1”

Software simulation
of external clearance

AAlx

Clears maximum storage

Clear Peak Memory
t

clear

AAlzX

X=0 bis 9
0=AUS

1=10

ms

2=50

ms

3=250

ms

4=1 s 5=5 s

6=25 s

7=EXTERN

8=AUTO

9=HOLD

Basic

range

(read)

AAmb

Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)

Sub

range

(read)

AAme

Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)

Sub range (set)

1. AAm1XXXXYYYY

XXXXYYYY=2x4 hex-digit for lower and upper sub range

limit (°C or °F)

2. AAm2

AAm2 confirms the change (auto reset)

Measuring

value

AAms

Answer: QQQQQ (88880=Overflow)
5 decimal digit (in °C or °F, last digit is 1/10 °C or °F)

Device type

AAna

Output: “IS 6 Advanced / IGA 6 Advanced”
(16 ASCII-characters)

Read

Parameters

AApa

Answer: 15 decimal digits
DD.............

: Emissivity (see em)

..D............

: Response time (see ez)

...D...........

: Clear peak memory (see lz)

....D..........

: Analog output (see as)

.....DD........

: Internal temperature (see gt)

.......DD......

: Device address (see ga)

.........D.....

: Baud rate (see br )

..........01000

: Fixed value

Serial number

AAsn

Output: XXXXX (hex 5-digit)

Advertising