IS 6 Advanced / IGA 6 Advanced Manual

Troubleshooting · 42

Symptom

Probable Cause

Comments

Temperature readings
are too high

Faulty IS 6 Advanced /
IGA 6 Advanced

Contact LumaSense. See
section

1.5

Electrical noise in lead wires caused
by intense magnetic fields and/or
improper selection of interconnecting
cable and/or cable run in same
conduit as AC power lines

See section

3.1

Lens or window clouding up

Increase or decrease air purge

rate and/or investigate other
cause.

Improper grounding of cable shield
and/or IS 6 Advanced /
IGA 6 Advanced housing (or cooling
jacket)
IS 6 Advanced / IGA 6 Advanced

mounting not secure

Noisy readings (fast
fluctuations)

Target moves more than 95% out of
field of view. Conveyor type
applications will also cause
fluctuations as targets move by the
field of view

Application may require
Maximum to be stored. Apply
t

clear

settings. See section

4.5

Target may be changing temperature

quickly. (Thermocouples do not
respond as fast as the IS 6 Advanced /
IGA 6 Advanced)

Speed of response may be

slowed down
Apply t

90

settings. See section

4.4

Target may be reflecting nearby

objects that have fluctuating
temperatures
Loose connection in wiring

Electrical noise lead wires caused by
intense magnetic fields, improper
selection of inter-connecting cable
Cable run into in same conduit as AC
power lines
Flame or reflections of flame may be

entering the field of view
Extremely dense steam, smoke, thick
dust, personnel or moving machinery
intermittently blocking 95% of
optical path of IS 6 Advanced /
IGA 6 Advanced
Non-uniform emissivity of large

moving targets

Slower response speed is

required

Faulty IS 6 Advanced /
IGA 6 Advanced

Contact LumaSense. See
section

1.5

Laser Targeting Light
fails

The instrument‘s internal
temperature is higher than 61 °C

Use cooling jacket