IS 6 Advanced / IGA 6 Advanced Manual

Data format UPP

Ò

· 37

Description

Command Parameters

Device type /
Software version

AAve

Answer: VVMMJJ VV=54
MM=Month JJ = Year of software version

Communication
Module/ software
version in detail

AAvc

tt.mm.jj XX.YY
tt = day; mm = month; jj = year;
XX.YY = Software version

Software version in

detail

AAvs

tt.mm.jj XX.YY

tt = day; mm = month; jj = year;
XX.YY = Software version

Note: the letter “l” means the lower case letter of “L”.

Additional instruction for the RS485 interface:

Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some

older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 5 ms, at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be

repeated.

4. After receiving the response, the master has to wait at least 1.5 ms before a new

command can be entered.