7.2 Error Messages and Responses
APS-1102 Programmable AC/DC Power Source
7-9
Table 7-5. Panel Operation Errors (2/2)
Error Message
Cause and Action Required
UNDER BUSY STATE
Cannot set (execute) due to BUSY status. Wait until the BUSY display
disappears, then try operation again.
(
See “3.4.3 Setting output voltage range”.)
UNDER REMOTE STATE
Cannot set (execute) due to remote status. Set local mode.
(
See “5.5.3 Remote”.)
UNDER LOCAL STATE
Cannot set (execute) due to local mode. (Remote mode only)
(
See “4.5 Control Using External Control I/O Connector”.)
(
See “6 USB INTERFACE”.)
UNDER KEYLOCK
Cannot set (execute) due to keylock status. Cancel keylock.
(
See “5.6.1 Keylock”.)
OUT OF LIMITER
Cannot set (execute) because range limit is exceeded. (Remote mode
only)
(
See “4.5 Control Using External Control I/O Connector”.)
(
See “6 USB INTERFACE”.)
COMMAND ERROR
This is the USB interface command error. Check the command.
(Remote mode only)
(
See “6.6 Error Message List”.)
EXECUTION ERROR
Execution errors other than the above.