2.4 Simple Operation Checks
APS-1102 Programmable AC/DC Power Source
2-9
2. Panel settings
Set the APS-1102 as shown below.
See “3.4 Main Operation Example”, for descriptors
of the settings methods.
Table 2-1. Panel Settings for Operation Check
Item Setting
Output mode
AC-INT mode
Output voltage range
100 V
AC output voltage
100 Vrms
Selected measurement display RMS
3. Press
the
OUTPUT
key to turn on output.
OUTPUT
OUTPUT
Output is off.
Output is on.
⇒
Check that the control panel MEASURE screen shows 100 Vrms as the AC output voltage.
Figure 2-3. Operation Check
⇒
If the normal values are not displayed, it may be due to an operation fault. Contact Good Will
or a Good Will representative.
4. Turn off the power, check that OUTPUT off is displayed, then press the power switch to the bottom
() position. This shuts off the power supply and turns off the power source unit.
Before turning off the power, make sure that the output on/off LED is off.
Output on/off LED is on.
Check display.
!
CAUTION