GW Instek APS-1102 User Manual User Manual
Page 99

4.4 Using the Sequence Function
APS-1102 Programmable AC/DC Power Source
4-35
Operation step <2>
The settings for DC voltage operation type (AC+DC mode only), AC voltage operation type,
frequency operation type, waveform (WAVE), waveform operation type, phase operation type, step
synchronized output (SYNC CODE), step synchronized output operation type, step end phase
enable/disable (STOP PHS), and step end (STEP TERM) are made by selecting settings for the various
items in the order described below.
The example described below is when entering the step end (STEP TERM) setting.
1. Open the SEQUENCE screen and display the PROGRAM screen.
See “4.4.2 Sequence settings”, for steps for displaying the PROGRAM screen.
2. Press the
or
key to move the cursor to “STEP TERM”.
ENTER
CANCEL
Cursor
3. Press the
ENTER
key to display the selection box.
ENTER
CANCEL
Selection box
4. Press the
or
key to move the cursor to the operation to be set.
ENTER
CANCEL
5. Press the
ENTER
key to set the operation at the step end.