3.4 Main Operation Example
APS-1102 Programmable AC/DC Power Source
3-19
3.4.3 Setting output voltage range
Select 100 V or 200 V. The output voltage range cannot be changed while output is on.
The setting range varies according to the output voltage range.
Table 3-5. List of Setting Ranges for Various Output Voltage Range
Setting
Setting Range
100 V range
200 V range
Unit
DC voltage
200.0 to +200.0
400.0 to +400.0
V
AC voltage
Wave-
form
SIN/SQU
0.0 to 140.0
0.0 to 280.0
Vrms
ARB1 to 16
0.0 to 400.0
0.0 to 800.0
Vp-p
Output peak current limiter (positive)
+10.0 to +42.0
+5.0 to +21.0
A
Output peak current limiter
(negative)
42.0 to 10.0
21.0 to 5.0
A
Output average current limiter
1.0 to 10.5
1.0 to 5.3
A
Positive voltage setting limit
+5.0 to +200.0
+10.0 to +400.0
V
Negative voltage setting limit
200.0 to 5.0
400.0 to 10.0
V
External input gain
0.0 to 200.0
0.0 to 400.0
times
Operation steps
To set the output voltage range, select
in the SET menu.
1. Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor