GW Instek APS-1102A User Manual User Manual
Page 209

6 REMOTE INTERFACE
APS-1102A
6-25
Fourth parameter (AC voltage operation type)
Type
<NR1>
Range
0 (CONST), 1 (KEEP), or 2 (SWEEP)
Fifth parameter (frequency)
Type
<NR2>
Range
1.0 to 550.0
Sixth parameter (frequency operation type)
Type
<NR1>
Range
0 (CONST), 1 (KEEP), or 2 (SWEEP)
Seventh parameter (waveform)
*The unit for the third parameter AC voltage varies depending on the seventh parameter setting.
Type
<NR1>
Range
0 (SIN), 1 (SQU), or 2 to 17 (ARB1 to ARB16)
Eighth parameter (waveform operation type)
Type
<NR1>
Range
0 (CONST) or 1 (KEEP)
Ninth parameter (start phase)
Type
<NR2>
Range
0.0 to 359.9
10th parameter (start phase operation)
Type
<NR1>
Range
0 (CONST) or 1 (KEEP)
11th parameter (step synchronized output)
Type
<NR1>
Range
0 (LL), 1 (LH), 2 (HL), or 3 (HH)
12th parameter (step synchronized output operation type)
Type
<NR1>
Range
0 (CONST) or 1 (KEEP)
Command example: SEQ:EPAR 0, 0, 100.0, 0, 50.0, 0, 1, 0, 90.0, 0, 0, 0
Query example:
SEQ:EPAR?
Response example:
0, 0, 100.0, 0, 50.0, 0, 1, 0, 90.0, 0, 0, 0