GW Instek APS-1102A User Manual User Manual
Page 64

APS-1102A User Manual
APS-1102A
3-30
b) Using setting range limiter
The APS-1102A internal mode (AC-INT, AC+DC-INT) and internal + external mode (AC-ADD,
AC+DC-ADD) include setting range limiter functions for output voltage and output frequency.
S
ee “5.5.5 Setting range limiter (LIMIT)”, for details.
Table3-13. Voltage and Frequency Setting Range Limiter
Setting
Setting Range
Resolution
Initial Value
Positive voltage
setting limit
100 V range
+0.1 V to +220.0 V
0.1 V
+220.0 V
200 V range
+0.1 V to +440.0 V
0.1 V
+440.0 V
Negative voltage
setting limit
100 V range
220.0 V to
0.1 V
0.1 V
220.0 V
200 V range
440.0 V to
0.1 V
0.1 V
440.0 V
Frequency upper limit setting limit
1.0 Hz to 550.0 Hz
0.1 Hz
550.0 Hz
Frequency lower limit setting limit
1.0 Hz to 550.0 Hz
0.1 Hz
1.0 Hz
Operation steps
The setting range limit is set via the LIMIT screen under the MISC menu
.
1. Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor
2. Press the
EN TER
key to display the MISC menu.
Press the
or
to move the cursor key to “LIMIT”.
ENTER
CANCEL