GPT-9000/9000A Series User Manual
128
Parameter
(MANU mode)
No parameter needed for MANU
mode.
Parameter
(AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter <string>
Returns the test status of the test
in the following format:
function, judgment or status, test
voltage, test current or resistance,
test time or ramp time
Function
ACW, DCW, IR, GB
Judgment
/Status
PASS, FAIL
VIEW
Test voltage
voltage+unit
Test current
/Test resistance
current+unit
resistance+unit
Test time
/Ramp time
T=time+S
R=time+S
Example
(in MANU mode)
MEAS?
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test.
Example
(in AUTO mode)
MEAS10?
>IR, FAIL ,0.225kV ,999M ohm,T=010.3S
Returns step 10 of the current automatic result.
MAIN:FUNCtion
Set
Query
Description
Changes the mode between AUTO and MANU.
Syntax
Query Syntax
MAIN:FUNCtion {MANU|AUTO}
MAIN:FUNCtion ?
Parameter/
Return parameter
MANU
Puts the tester mode to MANU.
AUTO
Puts the tester mode to AUTO.
Example
MAIN:FUNC MANU
Sets the tester to MANU mode.