Example
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.
MANU:DCW:REF
Set
Query
Description
Sets or returns the DCW reference value in mA.
The test must first be in DCW mode before this
command can be used.
The reference value must be less than the HI SET
value.
The reference value uses the same range as the HI
SET value.
Syntax
Query Syntax
MANU:DCW:REF <NR2>
MANU:DCW:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX/99XXA)
Example
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
MANU:DCW:ARCCurrent
Set
Query
Description
Sets or returns the DCW ARC current value in mA.
ARC must be enabled to set the ARC current. The
test must first be in DCW mode before this
command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
Syntax
Query Syntax
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/
Return parameter
<NR2>
1.000 ~ 20.00 (GPT-98XX)
2.000 ~ 040.0 (GPT-99XX/99XXA)