REMOTE CONTROL
131
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function
only.
Syntax
Query Syntax
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/
Return parameter
<NR2>
0.1~999.9 seconds
Example
MANU:RTIM 0.5
Sets the ramp time to half a second.
MANU:EDIT:MODE
Set
Query
Description
Sets or returns the mode (ACW, DCW, IR, GB) of
the selected manual test.
Syntax
Query Syntax
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?
Parameter/
Return parameter
ACW
AC Withstand mode
DCW
DC Withstand mode
IR
Insulation Resistance mode
GB
Ground Bond mode
Example
MANU:EDIT:MODE ACW
Sets the mode to ACW.
MANU:ACW:VOLTage
Set
Query
Description
Sets or returns the ACW voltage in kV. The test
must first be in ACW mode before this command
can be used.
Syntax
Query Syntax
MANU:ACW:VOLTage <NR2>
MANU:ACW:VOLTage?
Parameter/
Return parameter
<NR2>
0.100 ~ 5.000 (kV)
Example
MANU:ACW:VOLT 1
Sets the ACW voltage to 1 kV.