GW Instek GPT-9900 series User Manual User Manual
Page 79

OPERATION
79
Running the Test
1. In special test mode (000), tests are
started and stopped in the same
way as for the normal manual test
mode. See page 64 for details.
2. If required, the scroll wheel can be
used to set the voltage level in real-
time as the test is running (this
does not apply to IR or GB tests).
ACW
DCW
0.100kV ~ 5kV
0.100kV ~ 6kV
Results
Test judgments are the same as
those for the normal manual tests.
Please see the PASS/FAIL MANU
Test section for details.
View Sweep
Graph
Unlike normal manual tests, the special test
mode also has an option to view the resultant
test as a sweep graph.
This option is only available for the
GPT-9901A/9902A/9903/9903A/9904.
Steps
1. When the test has finished, press
the SWEEP key to view the results
of the sweep in a graph.
W E E
S
P
Graph Test Items:
TEST
Primary
Secondary
ACW
Test voltage
test current
DCW
Test voltage
test current
IR
Test voltage
test resistance
GB
Test current
test resistance