Setting a Reference Value
Background
The REF# acts as an offset. The REF# value is
subtracted from the measured current (ACW,
DCW) or measured resistance (IR, GB).
Steps
1. Press the UP / DOWN arrow keys
to bring the cursor to the
REF# setting.
I R
G B
m A
A CW
D CW
I / L
H
O
I M E
T
R
E
F R E Q =
0 H z
6
0
100
k
V
I
D
E
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
0
L
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
T
2
M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set the
REF# value.
ACW
DCW
IR
GB
0.000mA~HI SET current-0.1mA
0.000mA~HI SET current-0.1mA
0000MΩ~HI SETΩ-1MΩ
000.0mΩ~HI SETΩ-0.1mΩ
Note
For GB tests, a reference offset can be
automatically created using the zeroing function.
See page 73 for details.