GW Instek GPT-9900 series User Manual User Manual
Page 171

APPENDIX
171
Table 1a: Output Limitation in Withstanding Voltage Testing (GPT-98XX)
Upper Current Pause
Output Time
AC
30mA≤I≤40mA At least as long
as the output
time
Maximum 240
seconds
0.001mA≤I<30
mA
Not necessary Continuous
output possible
DC
0.001mA≤I≤10
mA
Not necessary Continuous
output possible
GB
15A<I≤30A
At least as long
as the output
time
999.9
3A≤I≤15A
Not necessary
999.9
NOTE: Output Time = Ramp Time + Test Time.
Table 1b: Output Limitation in Withstanding Voltage Testing (GPT-99XX/99XXA)
Upper Current Pause
Output Time
AC
80mA≤I≤100mA At least as long
as the output
time
Maximum 240
seconds
0.001mA≤I<80
mA
Not necessary Continuous
output possible
DC
0.001mA≤I≤20
mA
Not necessary Continuous
output possible
GB
15A<I≤32A
(99XX)
At least as long
as the output
time
999.9
3A≤I≤15A
Not necessary
999.9