PQA400 - PQA823 - PQA824

EN - 125

9. TECHNICAL SPECIFICATIONS

9.1. TECHNICAL

FEATURES

TRMS AC/DC Voltage Phase-Neutral / Phase-Ground – Single phase / Three phase systems

Range Accuracy Resolution

Input

impedance

0.0

 600.0V

(0.5%rdg+2dgt)

0.1V

10M

Max crest factor = 2
Voltage values < 2.0V are considered zero by meter
Meter is connectable with external VT with ratio selectable from 1

 3000

TRMS AC/DC Voltage Phase-Phase – Three phase systems

Range Accuracy

Resolution

Input

impedance

0.0

 1000.0V

(0.5%rdg+2dgt)

0.1V

10M

Max crest factor = 2
Voltage values < 2.0V are considered zero by meter
Meter is connectable with external VT with ratio selectable from 1

 3000

Voltage Anomalies Phase-Neutral – Single phase / Three phase 4-wire systems

Range

Voltage

Accuracy

Time Accuracy

(50Hz)

Voltage Resolution

Time Resolution

(50Hz)

0.0

 600.0V

(1.0%rdg+2dgt)

10ms

0.2V

10ms

Max crest factor = 2
Voltage values < 2.0V are considered zero by meter
Meter is connectable with external VT with ratio selectable from 1

 3000

Selectable threshold from

1% to 30%

Voltage Anomalies Phase-Phase – Three phase 3-wire systems

Range

Voltage

Accuracy

Time Accuracy

(50Hz)

Voltage Resolution

Time Resolution

(50Hz)

0.0

 1000.0V

(1.0%rdg+2dgt)

10ms

0.2V

10ms

Max crest factor = 2
Voltage values < 2.0V are considered zero by meter
Meter is connectable with external VT with ratio selectable from 1

 3000

Selectable threshold from

1% to 30%

Voltage Spikes – Phase-Ground - Single phase and Three phase systems (only PQA824)

Range

Voltage

Accuracy

Voltage

Resolution

Time resolution

(50Hz)

Detection time (50Hz)

-1000V

 100V

(2.0%rdg+60V)

1V

10ms

78

s – 2.5ms

(SLOW)

100V

 1000V

-6000V

 -100V

(10%rdg+100V)

15V

5

s – 160s

(FAST)

100V

 6000V

Voltage detection threshold set from 100V

 5000V

Maximum number of recorded events: 20000

TRMS AC Current with Standard STD transducer clamp

Range Accuracy

Resolution

Input

impedance Overload

protection

0.0

1000.0mV

(0.5%rdg+0.06%FS)

0.1mV

510k

5V

Measure performed with voltage output =1VAC with nominal current measured by clamp
Max crest factor = 3
Current values < 0.1% of full scale (FS) are considered zero by meter

TRMS AC Current with Flex transducer clamp – Range 300A

Range

Accuracy Resolution

Input

impedance Overload

protection

0.0

 49.9A

(0.5%rdg+0.24%FS)

0.1A

510k

5V

50.0

 300.0A

(0.5%rdg+0.06%FS)

Measurements performed with HTFLEX33 flexible clamp
Max crest factor = 3
Current values < 1A are considered zero by meter

TRMS AC Current with Flex transducer clamp – Range 3000A

Range (*)

Accuracy Resolution

Input

impedance Overload

protection

0.0

 3000.0A

(0.5%rdg+0.06%FS)

0.1A

510k

5V

Measurements performed with HTFLEX33 flexible clamp
Max crest factor = 3
Current values < 5A are considered zero by meter