HT instruments PQA824 User Manual

Page 42

Advertising
background image

PQA400 - PQA823 - PQA824

EN - 41

5.3.1.1. Setting system frequency

1. Move the cursor using arrows keys on field relative to “Freq[Hz]”, marked with blue

background.

2. Use F3 or F4 keys (alternatively touch MOD(+) o MOD(-)) to select the system

frequency choosing between 50Hz or 60Hz options. This parameter is relevant ONLY if
the input voltage not permits to recognize the frequency value (for example, only the
clamps for the current measurement are connected). In this case the instrument
generates an internal synchronism equal to the value of the set frequency.

3. Press SAVE or ENTER keys (or the smart icon

) to save the selected option and

confirm by “Ok”. This selected parameter will be present also after turn off the meter.

4. Press ESC key (or the smart icon ) to quit without saving any modification.

5.3.1.2. Setting clamp type

This parameter must be always set equal to the clamp type used.
Two types of clamps are available:

 STD:

For Standard clamps or Current Transformer.

 FLEX:

For Flexible clamps.

1. Move the cursor using arrows keys on field relative to “Clamp Type”, marked with blue

background.

2. Use F3 or F4 keys (alternatively touch MOD(+) o MOD(-)) to select clamp type

choosing between STD or FLEX options.

3. Press SAVE or ENTER keys (or the smart icon

) to save the selected option and

confirm by “Ok”. This selected parameter will be present also after turn off the meter.

4. Pressing ESC key (or the smart icon ) to quit without saving any modification.

5.3.1.3. Setting full scale of clamps

The value of this parameter must be always equal to the full scale of the current
clamps
used to take the measurement. In case multi-scale clamps are used, the value of
this parameter must be equal to the scale selected on the clamps.

1. Move the cursor using arrows keys on field relative to “Clamps FS [A]” , marked it with

blue background.

2. Use F3 or F4 keys (alternatively touch MOD(+) o MOD(-)) to select the desired full

scale. In case of STD clamps any value can be set by using F3 or F4 keys (alternatively
touch MOD(+) o MOD(-)). In case of FLEX clamp only 300A or 3000A options are
possible.

3. Press SAVE or ENTER keys (or the smart icon

) to save the selected option and

confirm by “Ok”. This selected parameter will be present also after turning off the meter.

4. Pressing ESC key (or the smart icon ) to quit without saving any modification.