Jenway 3510 User Manual

Page 12

Advertising
background image

351 050/REV B/09-06

9

3.6 Instrument Set-Up

The following section details the set-up modes available to the user.
These are: Number of calibration points entry, Buffer type entry, Manual buffer entry 1-3, Display
resolution, Manual temperature, Temperature units and serial port Baud Rate. These can be set in
sequence as detailed in this section or, by entering the SETUP mode and using the STO key to
select a specific parameter at any time.

To exit the set up menus at any time press the ESC key. This will return the instrument to the
MODE menu. Any parameters not saved will remain as defaults or previous setting.

3.6.1 Number of calibration points entry

I, 2 or 3 point calibration is possible on the 3510. To set these parameters:

Select SETUP mode on the display using the Left arrow key.

The secondary display will show CALPTS and will then scroll CALIBRATION POINTS after 10
seconds.

Select the required number of calibration points using the Up/Down arrow keys. The Left/Right
arrow keys and the CAL/CLR key have no function during this set up.

Symbol will flash while adjustment is being made.

To accept the number of calibration points entered, press the STO key. The symbol will disappear
and the secondary display will momentarily show SAVED and the next set up menu option will be
displayed.

Advertising