Jenway 3510 User Manual

Page 14

Advertising
background image

3.6.3 pH resolution

Note:

Setting the pH resolution will affect the resolution on the main instrument display.
Manual buffer entry is always shown to 3 decimal places.

The display will show:

The secondary display will show PH RES and will then scroll PH RESOLUTION after 10 seconds.

Select the preferred resolution from the options (0.001, 0.01 or 0.1) using the Up/Down arrows.

The Left/Right arrow keys and the CAL/CLR key have no function during this set up.

Symbol will flash while adjustment is being made.

To accept the selected resolution, press the STO key. The symbol will disappear and the secondary
display will momentarily show SAVED and the next set up menu option will be displayed.

To exit the set up menus at any time press the ESC key. This will return the instrument to the
MODE menu.

11

351 050/REV B/09-06

The correct values should be entered and saved as for the 1 point calibration. When all chosen
buffer types have been entered and confirmed the instrument display will update to show the next
set up menu option.

To exit the set up menus at any time press the ESC key. This will return the instrument to the
MODE menu.

If a 2 or 3 point calibration is being performed the display will move on to the 2 and then 3 point
calibration buffer set up screens.