Interfacing (continued)

T

he RS232 communications parameters on the computer or printer need to be set to match those of

the Model 3510, as detailed below:

1200 Baud

9600 Baud

7 Data Bits

8 data bits

Odd Parity

OR

No parity

1 Stop Bit

1 stop bit

Setting of these options is detailed in Section 3.6.6.

The Model 3510 supports both hardware (CTS/RTS) flow control and software XON/XOFF flow control.

Pressing the PRINT key outputs from the RS232 interface.

Sending an ASCII “D” to the 3510 causes a printout of the current displayed reading plus sample
number.

Sending an ASCII “C” causes a printout of the last calibration parameters.

Sending an ASCII "P" causes a printout of the stored readings.

6.3 Keypad Emulation

351 050/REV B/09-06

24

Keypad remote control using RS232 interface:

7

- Instrument On / Standby / Escape

1

- Calibrate / Memory Clear

9

- Print

8

- Up Arrow

2

- Down Arrow

4

- Left Arrow

6

- Right Arrow

3 or 5

- Enter / Store