1.1 Instrument Description

The Model 3510 is a general purpose pH/mV/Temperature bench meter used for routine laboratory
analysis. The meter supports 1, 2 or 3 point pH calibration on either manually entered pH buffer
values or automatically temperature compensated buffers to DIN, JIS and NIST standards and
Jenway buffers supplied with the instrument. Up to 3 decimal place resolution is available. The 3510
includes a 32 reading memory facility.

1.2 Instrument Specification

pH (1, 2 or 3 point cal)
Range:

-2.000 to 19.999pH

Resolution:

0.001 / 0.01 / 0.1pH

Accuracy:

±0.003pH

mV (Absolute or Relative)
Range:

-1999 to +1999mV

Resolution:

0.1mV

Accuracy:

±0.2mV

Input Impedance:

>10

12

ohms

Temperature Measuring
Ranges:

-10 to +105°C / 14 to 221°F

Resolution:

0.1°C / 1°F

Accuracy:

±0.5°C / ±1°F

ATC Range:

0 to 100°C / 32 to 212°F

Manual Temp. Compensation:

0 to 100°C / 32 to 212°F

Auto Buffer Selection:

Jenway (2.00, 4.00, 7.00, 9.20 and 10.00)
DIN (3.06, 4.65, 6.79, 9.23, 12.75)
NIST (1.68, 4.01, 6.87, 9.18, 12.45)
JIS (1.68, 4.01, 6.87, 9.18, 12.45) or manually entered buffers

Calibration:

User selectable 1, 2 or 3 point

Outputs:

Analogue 1mV per 0.01pH
RS232

Display:

Back lit custom LCD

Power:

Power Supply 9Vac

Size:

275(l)x240(w)x150(d)mm

Weight:

850g